Battery high and low temperature low pressure environment test chamber temperature and humidity test box simulation low pressure test chamber

Product Usage: The low temperature and low pressure test chamber mainly provides temperature and humidity change environment for scientific research and production units such as aerospace, aviation, petroleum, chemical industry, automobile, ship, electronics, communication, etc. Provide reliability testing, product screening, etc. for the safety testing of electronic components. It is widely used in high temperature test, low temperature test, temperature-low pressure test function of various connectors, cables, micro-motors, etc. Fulfill the standard: 3.1.GB/T 2423.1-2008 Test A: Low temperature test method 3.2. GB/T 2423.2-2008 Test B: High Temperature Test Method 3.3. GB/T2423.3-2008 Constant Damp Heat Test 3.4. GB /T2423.42008 Alternating Damp Heat Test 3.5. GB /T2423.21-2008 Low Pressure Test Method 3.6.GB/T2423.22-2008 test Nb temperature change test 3.7.GB/T2423.25-2008 test Z/AM: low temperature/low pressure comprehensive test; 3.8. GB /T2423.26-2008 test Z / BM; high temperature / low pressure comprehensive test; 3.9.GJB150.2A-2009 Low pressure (altitude) test and other test standards
technical parameter  
temperature range 60℃~+150℃
temperature fluctuation ≤±0.5℃ (at normal pressure, no load, constant state)
temperature uniformity ≤2℃ (at normal pressure, no load, constant state)
temperature deviation ≤±2℃ (at normal pressure, no load, constant state)
Heating and cooling rate Air temperature: -60℃-+150℃, the whole process average ≥1℃/min, normal  pressure, with load of 10kg aluminum ingot;
  Air temperature: +150℃-60℃, the whole process average ≥1℃/min,  normal pressure, with a load of 10g aluminum ingot
pressure range normal pressure~0.5kPa (common pressure 62kPa, simulated altitude 4km)
buck rate normal pressure~1.0kPa depressurization time≤30min
  pressure recovery rate: ≤10.0kPa/min
fast buck rapid decompression: 75.2kPa--18.8 kPa ≤15s
pressure deviation when the pressure is ≥40KPa: ≤±2KPa
  when the pressure is 4KPa~40KPa: ≤±5%
  when the pressure is ≤4KPa: ≤±0.1KPa
rate of change of air pressure K15kPa/min (adjustable)
fulfill the standard CB/T 2423.212008 Basic environmental test procedures for electrical and electronic  products Test method for low air pressure
acceptance Criteria CB/T5170
CB/T5170

  • Battery high and low temperature low pressure environment test chamber temperature and humidity test box simulation low pressure test chamber Related Video: